Sealed Micro-Electrochemical H-Cell Setup
Captions
- H-type Sealed Micro-Electrochemical Cells: 10/20 ml H-cell (C2032G1) and 5/15 ml H-cell (C2032G1)
- Three Electrode Slots and Two Gas Inlets/ Outlets included
- Let’s quickly see how to assemble
- Use membrane sealer to fix your membrane in between the compartments
- Use a membrane separator suitable for the electrolyte
- Tighten the junction against leak using the provided clamp (Electrodes needs to purchase separately)
- Working Electrode: You can use an Electrode Holders or a Disk Electrode with the Cell
- Disk Working Electrode fixing (Platinum and Glassy Carbon Normal and Removable Disk WE are available to purchase alongside Platinum and Glassy Carbon Electrode Holders)
- Reference Electrode fixing
- Ag/AgCl for Aqueous and Ag/Ag+ for non-aqueous
- Allow the Reference Electrode the reach the equilibrium once fills with the solution before use
- Always makes sure to use the provided O-rings to achieve adequate sealing
- Counter Electrode fixing
- Wire, and Coil Counter/ Auxiliary Electrodes fits well with the cell
- Can fix the Gas Inlet/Outlets or close the ports with Stoppers (provided) as required
- Follow the steps to obtain good sealing
- Now your cell is ready
- Around 3ml minimum Electrolyte in each compartment is needed to immerse the Electrodes
- Electrodes are well immersed with 5 ml Electrolyte in each compartment
- The cell has maximum 10-12 ml capacity in each compartment
- A range of spare parts and extras and included within:
-
- O-rings for PTFE Lids
- PTFE Tubing (1m)
- PTFE Sealing
- Electrode and Gas Feed Fitting Screws and O-rings
- PTFE Rod Stoppers and Caps
- Silicone Rubber Gaskets
For further assistance, contact us via the product page or at support@ossila.com