LED Measurement System User Manual
Please select the correct version of the user manual for your LED Measurement System. You can find the model number on the sticker on the base of your unit.
LED Measurement System
1. Overview
The Ossila LED Measurement System is a low-cost solution for reliable current-voltage-luminance measurements of light emitting diodes. The system is controlled by specially designed software which can perform multiple IVL measurements and measure the current, luminance, and efficiency over long periods of time.
2. EU Declaration of Conformity (DoC)
We
Company Name: Ossila BV
Postal Address: Biopartner 3 Building, Galileiweg 8
Postcode: 2333 BD Leiden
Country: The Netherlands
Telephone Number: +31 (0)718 081020
Email Address: info@ossila.com
declare that the DoC is issued under our sole responsibility and belongs to the following product:
Product: Ossila LED Measurement System (T2005A3/T2005B3/T2005C3/T2005E3)
Serial Number: T2005A3 - xxxx, T2005B3 - xxxx, T2005C3 - xxxx, T2005E3 - xxxx
Object of Declaration
Ossila LED Measurement System (T2005A3/T2005B3/T2005C3/T2005E3)
The object of declaration described above is in conformity with the relevant Union harmonisation legislation:
EMC Directive 2014/30/EU
RoHS Directive 2011/65/EU
Signed:
Name: Dr James Kingsley
Place: Leiden
Date: 16/11/2021
Декларация | за съответствие на ЕС |
---|---|
Производител | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Декларира с цялата си отговорност, че посоченото оборудване съответства на приложимото законодателство на ЕС за хармонизиране, посочено на предходната(-ите) страница(-и) на настоящия документ. |
[Čeština] | Prohlášení o shodě EU |
---|---|
Výrobce | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Prohlašujeme na vlastní odpovědnost, že uvedené zařízeni je v souladu s příslušnými harmonizačními předpisy EU uvedenými na předchozích stranách tohoto dokumentu. |
[Dansk] | EU-overensstemme lseserklærin g |
---|---|
Producent | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Erklærer herved, at vi alene er ansvarlige for, at det nævnte udstyr er i overensstemmelse med den relevante EUharmoniseringslovgivning, der er anført på den/de foregående side(r) i dette dokument. |
[Deutsch] | EU-Konformitätserklärung |
---|---|
Hersteller | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Wir erklären in alleiniger Verantwortung, dass das aufgeführte Gerät konform mit der relevanten EUHarmonisierungsgesetzgebung auf den vorangegangenen Seiten dieses Dokuments ist. |
[Eesti keel] | ELi vastavusavaldus |
---|---|
Tootja | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Kinnitame oma ainuvastutusel, et loetletud seadmed on kooskõlas antud dokumendi eelmisel lehelküljel / eelmistel lehekülgedel ära toodud asjaomaste ELi ühtlustamise õigusaktidega. |
[Ελληνικά] | Δήλωση πιστότητας ΕΕ |
---|---|
Κατασκευαστής | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Δηλώνουμε υπεύθυνα όn ο αναφερόμενος εξοπλισμός συμμορφώνεται με τη σχεnκή νομοθεσία εναρμόνισης της ΕΕ που υπάρχει σnς προηγούμενες σελίδες του παρόντος εγγράφου. |
[Español] | Declaración de conformidad UE |
---|---|
Fabricante | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Declaramos bajo nuestra única responsabilidad que el siguiente producto se ajusta a la pertinente legislación de armonización de la UE enumerada en las páginas anteriores de este documento. |
[Français] | Déclaration de conformité UE |
---|---|
Fabricant | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Déclarons sous notre seule responsabilité que le matériel mentionné est conforme à la législation en vigueur de l'UE présentée sur la/les page(s) précédente(s) de ce document. |
[Hrvatski] | E.U izjava o sukladnosti |
---|---|
Proizvođač | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Izjavljujemo na vlastitu odgovornost da je navedena oprema sukladna s mjerodavnim zakonodavstvom EU-a o usklađivanju koje je navedeno na prethodnoj(nim) stranici(ama) ovoga dokumenta. |
[Italiano] | Dichiarazione di conformità UE |
---|---|
Produttore | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Si dichiara sotto la propria personale responsabilità che l'apparecchiatura in elenco è conforme alla normativa di armonizzazione UE rilevante indicata nelle pagine precedenti del presente documento. |
[Latviešu] | ES atbils tības deklarācija |
---|---|
Ražotājs | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Ar pilnu atbilclību paziņojam, ka uzskaitītais aprīkojums atbilst attiecīgajiem ES saskaņošanas tiesību aktiem, kas minēti iepriekšējās šī dokumenta lapās. |
[Lietuvių k.] | ES atitikties deklaracija |
---|---|
Gamintojas | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
atsakingai pareiškia, kad išvardinta įranga atitinka aktualius ES harmonizavimo teisės aktus, nurodytus ankstesniuose šio dokumento |
[Magyar] | EU-s megfelelőségi nyilatkozat |
---|---|
Gyártó | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Kizárólagos felelösségünk mellett kijelentjük, hogy a felsorolt eszköz megfelel az ezen dokumentum előző oldalán/oldalain található EU-s összehangolt jogszabályok vonatkozó rendelkezéseinek. |
[Nederlands] | EU-Conformiteitsverklaring |
---|---|
Fabrikant | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Verklaart onder onze uitsluitende verantwoordelijkheid dat de vermelde apparatuur in overeenstemming is met de relevante harmonisatiewetgeving van de EU op de vorige pagina('s) van dit document. |
[Norsk] | EU-samsvarserklæ ring |
---|---|
Produsent | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Erklærer under vårt eneansvar at utstyret oppført er i overholdelse med relevant EU-harmoniseringslavverk som står på de(n) forrige siden(e) i dette dokumentet. |
[Polski] | Deklaracja zgodności Unii Europejskiej |
---|---|
Producent | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Oświadczamy na własną odpowiedzialność, że podane urządzenie jest zgodne ze stosownymi przepisami harmonizacyjnymi Unii Europejskiej, które przedstawiono na poprzednich stronach niniejszego dokumentu. |
[Por tuguês] | Declaração de Conformidade UE |
---|---|
Fabricante | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Declara sob sua exclusiva responsabilidade que o equipamento indicado está em conformidade com a legislação de harmonização relevante da UE mencionada na(s) página(s) anterior(es) deste documento. |
[Română] | Declaraţie de conformitate UE |
---|---|
Producător | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Declară pe proprie răspundere că echipamentul prezentat este în conformitate cu prevederile legislaţiei UE de armonizare aplicabile prezentate la pagina/paginile anterioare a/ale acestui document. |
[Slovensky] | Vyhlásenie o zhode pre EÚ |
---|---|
Výrobca | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Na vlastnú zodpovednosť prehlasuje, že uvedené zariadenie je v súlade s príslušnými právnymi predpismi EÚ o harmonizácii uvedenými na predchádzajúcich stranách tohto dokumentu. |
[Slovenščina] | Izjava EU o skladnosti |
---|---|
Proizvajalec | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
s polno odgovornostjo izjavlja, da je navedena oprema skladna z veljavno uskladitveno zakonodajo EU, navedeno na prejšnji strani/prejšnjih straneh tega dokumenta. |
[Suomi] | EU-vaatimustenm ukaisuusvakuutus |
---|---|
Valmistaja | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Vakuutamme täten olevamme yksin vastuussa siitä, että tässä asiakirjassa luetellut laitteet ovat tämän asiakirjan sivuilla edellisillä sivuilla kuvattujen olennaisten yhdenmukaistamista koskevien EU-säädösten vaatimusten mukaisia. |
[Svenska] | EU-försäkran om överensstämmelse |
---|---|
Tillverkare | Ossila BV, Biopartner 3 building, Galileiweg 8, 2333 BD Leiden, NL. |
Vi intygar härmed att den utrustning som förtecknas överensstämmer med relevanta förordningar gällande EUharmonisering som fmns på föregående |
3. Safety
3.1 Use of Equipment
The Ossila LED Measurement System is designed to be used as instructed. It is intended for use under the following conditions:
- Indoors in a laboratory environment (pollution degree 2).
- Altitudes up to 2000 m.
- Temperatures of 5°C to 40°C; maximum relative humidity of 80% up to 31°C.
The unit is supplied with a 24 VDC power adapter, in accordance with European Commission regulations and British Standards. Use of any other electrical power cables, adaptors, or transformers is not recommended.
3.2 Hazard Icons
The following symbols can be found at points throughout the rest of the manual. Note and read each warning before attempting any associated operations associated with it:
Symbol | Associated Hazard |
---|---|
Electrical shock |
3.3 General Hazards
Before installing or operating the Ossila LED Measurement System there are several health and safety precautions which must be followed and executed to ensure safe installation and operation.
3.4 Power Cord Safety
Emergency power disconnect options: use the power cord as a disconnecting method and remove from wall. To facilitate disconnect, make sure the power outlet for this cord is readily accessible to the operator.
3.5 Servicing
If servicing is required, please return the unit to Ossila Ltd. The warranty will be invalidated if:
- Modification or service has been carried out by anyone other than an Ossila engineer.
- The Unit has been subjected to chemical damage through improper use.
- The Unit has been operated outside the usage parameters stated in the user documentation associated with the Unit.
- The Unit has been rendered inoperable through accident, misuse, contamination, improper maintenance, modification, or other external causes.
3.6 Health and Safety - Servicing
Servicing should only be performed by an Ossila engineer. Any modification or alteration may damage the equipment, cause injury, or death. It will also void your equipment’s warranty.
4. Requirements
Table 4.1 details the power requirements for the Ossila LED Measurement System, and the minimum computer specifications for the Ossila LED IVL software.
Power | 24 VDC |
---|---|
Operating Systems | Windows 10 or 11 (32-bit or 64-bit) |
CPU | Dual Core 2 GHz |
RAM | 2 GB |
Available Hard Disk Space | 121 MB |
Monitor Resolution | 1440 x 960 |
Connectivity |
USB 2.0 Ethernet (requires DHCP) |
5. Unpacking
5.1 Packing List
The standard items included with the Ossila LED Measurement System are:
- The Ossila LED Measurement System.
- 24 VDC power adapter.
- USB-B cable.
- USB memory stick pre-loaded with the user manual, USB drivers, quality control data, and software installer.
- Resistor test device.
5.2 Damage Inspection
Examine the components for evidence of shipping damage. If damage has occurred, please contact Ossila directly for further action. The shipping packaging will come with a shock indicator to show if there has been any mishandling of the package during transportation.
6. Specifications
The Ossila LED Measurement System specifications are shown in Table 6.1.
Voltage range | ±10 mV to ±10 V |
---|---|
Voltage accuracy | ±10 mV offset |
Voltage resolution | 170 μV |
Current range | ±10 nA to ±200 mA (5 ranges) |
Current accuracy | ±10 nA (at 20 μA range) |
Current resolution | 0.1 nA (at 20 μA range) |
Lux range | 0 to 100 klx (3 ranges) |
Lux accuracy | ±10% |
Lux resolution | 0.0144 (at 500 lx range) |
Substrate Size |
20 mm x 15 mm 25 mm x 25 mm |
Substrate Compatibility |
T2005B - ITO Glass Substrates (S211) |
Overall Dimensions |
Width: 151 mm Height: 50 mm Depth: 300 mm |
7. System Components
The Ossila LED Measurement System is comprised of three items: the Ossila LED Measurement System (Figure 7.1), and the Ossila LED IVL Software (Figure 7.2).
8. Installation
- Install the Ossila LED IVL software on your PC.
- Run the file ‘Ossila-LED-IVL-Installer-X-X-X-X.exe’ on the USB memory stick provided.
- Follow the on-screen instructions to install the software.
- Connect the 24 VDC power adaptor to the power socket on the rear of the unit.
- Connect the unit to your PC using the provided USB-B cable, or an Ethernet cable if preferred.
If the unit is not detected, or driver installation fails, please refer to the SMU USB Driver Installation Guide found on the USB memory stick.
9. Operation
9.1 Measurement Types
The Ossila LED IVL software can perform 2 different types of measurements. Each measurement type can be selected using the tabs at the top of the window. The available measurements are:
- Characterisation (Section 9.1.1).
- Lifetime (Section 9.1.2).
Each measurement type requires several settings to be selected before it can be performed. Settings that are shared between all measurements are detailed in Section 9.3. Measurement specific settings are detailed in Sections 9.3.3, and 9.3.4
9.1.1 Characterisation
The Characterisation tab is used to perform I-V measurements of LEDs, whilst measuring the illuminance, luminance, white count, and current and power efficiencies using the built-in light sensors.
The Lifetime tab is used to perform long-term measurements of LEDs by holding them at a specified voltage and periodically measuring the current of the LED and response of the light sensors.
9.2 Quick Start Guide
- Start the Ossila LED IVL software. The window shown in Figure 9.1 will open.
- Choose a measurement type as described in Section 9.1.
- Place your device in the device holder and close the lid.
- Set the appropriate settings in the software (explained in more detail in Section 9.3).
- Click the ‘Measure’ button.
- For each pixel, measurements are performed using the chosen measurement settings.
- Data is plotted as it is measured, using a solid line for primary (left) y-axis and a dashed line for the secondary (right) y-axis.
- This process is repeated until all pixels have been measured.
- If automatic saving is turned on, the measurement data and settings will then be saved.
9.3 Software Settings
There are several settings in the program which must be filled in before taking a measurement. These are found in the column on the left of the window, as shown in Figure 9.1.
9.3.1 System Settings
(I) System Address
Select the COM port or IP address of the connected unit you intend to use (USB and Ethernet connection respectively). This box will be populated automatically with the addresses of any units connected to the computer or network.
(II) Current Range
Select the range of current values to be used for the measurement. This defines the upper limit and accuracy of current measurements that can be performed by the unit. The values for each range are given in Table 9.1.
Maximum Current | Accuracy | Precision | Resolution |
---|---|---|---|
±200 mA | ±500 μA | 10 μA | 1 μA |
±20 mA | ±10 μA | 1 μA | 100 nA |
±2000 μA | ±1 μA | 100 nA | 10 nA |
±200 μA | ±100 nA | 10 nA | 1 nA |
±20 μA | ±10 nA | 1 nA | 100 pA |
(III) Lux Range
Select the range of lux values to be used for the measurement. This defines the upper limit and resolution of lux measurements that can be performed by the unit. The values for each range are given in Table 9.2.
Maximum Lux | Accuracy | Resolution |
---|---|---|
100 klx | ±10% | 1.8432 lx |
5000 lx | ±10% | 0.1152 lx |
500 lx | ±10% | 0.0144 lx |
(IV) Sampling Rate
Select the number of samples to be taken for each data point. A higher number of samples per point will improve the accuracy and precision of the measurement. However, this will increase the time taken for the measurement to be performed.
9.3.2 Device Settings
(I) Pixels to Measure
Select which pixels to measure. Pixel numbers are labelled on the device holder. Selection boxes will be enabled and disabled based upon the device holder being used.
(II) Pixel Area
Set the area in cm2 of each pixel of the device.
(III) Inverted Device
Set whether the device to be measured is inverted. This option should be on if the anode of your device connects to the ‘cathode’ pins in the device holder.
9.3.3 Characterisation Settings
(I) Measure Luminance
Set whether measurements of the device’s light output will be performed. If set to ‘Off’, the system will not check that the lid is closed before measuring.
(II) Start Voltage
Set the voltage in volts at which to start the current-voltage measurement. This can be set between -10 V and +10 V.
(III) End Voltage
Set the voltage in volts at which to end the current-voltage measurement. This can be set between -10 V and +10 V.
(IV) Voltage Increment
Set the step size in volts for changing the voltage during current-voltage measurement.
(V) Voltage Settle Time
Set the time in seconds between applying a voltage and measuring the current. The minimum value is dependant on the selected lux range:
- 0 s for 100 klx
- 0.05 s for 5000 lx
- 0.4 s for 500 lx
(VI) Hysteresis I-V
This option performs a reverse current-voltage measurement after the forward currentvoltage measurement has completed. This reverses the set start and end voltages and uses the same voltage increment and settle time as the forward measurement.
9.3.4 Lifetime Settings
(I) Duration
Set the total duration in hours for the lifetime measurement.
(II) Interval
Set the time interval in minutes between performing repeat measurements of the device.
(III) Hold Voltage
Set the voltage that all pixels will be held at between measurements. This can be set between -10 V and +10 V.
(IV) Primary Axis Units
Set which property is plotted on the primary (left) y-axis.
(V) Secondary Axis Units
Set which property is plotted on the secondary (right) y-axis.
9.3.5 Saving and Loading Settings
(I) Save Settings
Saves the current settings as a profile that can be loaded quickly for use at another time. When clicked, you will be prompted to name the settings profile.
- If the name is already in use, you will be asked if you wish to overwrite the previous profile.
- The name cannot contain the characters: \ / : * ? “ < > |
The settings profile will be added to the drop-down box using the given name.
(II) Settings Profile
Select a saved settings profile from the drop-down box. The settings fields will be populated with the saved values from the selected profile.
Settings profiles can be deleted by selecting the profile and then clicking the red ‘delete’ icon next to the drop-down box.
9.3.6 Measurement Controls
(I) Measure
Clicking this button will start the measurement using the chosen settings. This button cannot be clicked if the software has not detected a unit.
(II) Abort
Stops a measurement that is currently in progress.
9.3.7 Plot Controls
(I) Plot Display Controls
By default, the axes of the plot will automatically scale to display all the data within it. The view can be controlled manually using the following mouse controls:
- Left/Middle click and drag – pan the axes.
- Right click and drag – scale the axes (left-right for x-axis, up-down for y-axis).
- Scroll wheel – scale the axes centred on the cursor location.
A specific axis can be controlled by using these controls on the axis labels. The axes can be reset by clicking the “Reset View” button shown in Figure 9.9.
(II) Axis Selection
In the characterisation tab the data plotted on the primary and secondary y-axes can be selected using the drop-down boxes beneath the axes. Axes cannot be changed whilst a measurement is being performed.
(III) Show Primary Axis
Controls whether the data plotted on the primary (left) y-axis is displayed.
(IV) Show Secondary Axis
Controls whether the data plotted on the secondary (right) y-axis is displayed.
9.3.8 Saving Data
(I) Save After Measurement
The program allows for data to be saved automatically, as well as manually once the measurement is complete. For automatic saving, the ‘Save Directory’ and ‘Device Label’ fields must be filled in before the measurement can start, these are detailed below.
(II) Save Directory
Sets the location in which to save the results. This can be set either by:
- Manually typing the directory into the field.
- Copying and pasting it from your file explorer.
- Clicking the folder icon in the field, which will open a dialog box to allow the selection of a folder to save to.
(III) Device Label
Sets the name of the comma-separated values (.csv) files in which data will be saved.
The field cannot contain the following characters: \ / : * ? “ < > |
(IV) Save Data
Clicking this button will manually save the measurement results of the active curve.
(V) Save Data Format
Data for each pixel is saved to a separate .csv file, with the following columns:
- Time (s) - (lifetime only)
- Voltage (V)
- Current (A)
- Illuminance (lx)
- Luminance (cd/m2)
- White Count
- Current Efficiency (cd/A)
- Power Efficiency (lm/W)
The settings used for a measurement are saved to a separate .csv file in a ‘Settings’ folder in the selected save directory. If this folder does not exist, it will be created.
10. Test Device
The system is shipped with a pair of test devices that can be used to verify the correct operation of the system. One has 100 kΩ resistors that can be used to check the calibration of the system. The other has semiconductor LEDs to test the response of the light sensors. Both are arranged in the geometry of the substrate pixels, and the appearance of the test device will depend on the substrate system being used.
10.1 Taking a Measurement
- Plug in and switch on the system.
- Allow at least 30 minutes for the system to warm up.
- Place the test device in the device holder with the resistors facing upwards.
- Start the LED IVL software and enter the following settings in Figure 10.2.
- These settings can be used with any current range except for the 20 μA range. For this range the start and end voltages must be lowered to -2 V and 2 V respectively.
- The ‘Pixels to Measure’ checkboxes should match the device configuration you have.
- Click the ‘Measure’ button.
- The system should measure straight line resistor responses from -100 μA to 100 μA (or - 20 μA to 20 μA for the 20 μA current range) as shown in Figure 10.3.
- To check the calibration of the system, use the I-V data to calculate the measured resistance at -10 and 10 V (-2 and 2 V for the 20 μA current range).
- Resistance can be calculated using: R = V / I
- For the 200 mA current range the calculated resistance should be between 98 and 102 kΩ (within 2% of the resistor value).
- For all other ranges the calculated resistance should be between 99 and 101 kΩ (within 1% of the resistor value).
11. Troubleshooting
Most of the issues that may arise will be detailed here. However, if you encounter any issues that are not detailed here, then please contact Ossila.
11.1 Installation and Setup
Problem | Possible Cause | Action |
---|---|---|
No power/display | The power supply may not be connected properly. | Ensure the system is firmly plugged into the power supply, and that the plug is connected to both the adaptor and a working power socket. |
The power supply adaptor has a fault. | Contact Ossila for a replacement power supply adaptor. | |
Software does not start | The wrong version of Windows is installed on the computer. | Install the software on a computer with Windows Vista or newer. |
The software has not installed properly. | Try reinstalling the software. | |
Cannot connect to the system via USB | The USB cable may not be connected properly. | Ensure the USB cable is firmly plugged in at both ends. |
The USB cable may not be connected to a working USB port. | Try connecting the unit to a different USB port on the computer. | |
The USB cable is defective. | Try using a different USB-B cable, and contact Ossila if necessary. | |
Cannot connect to the system via network | The MAC address of the unit is not registered with the internal network. | Register the system on the network using the MAC address obtained via a USB connection (see Source Measure Unit manual). |
The Ethernet cable may not be connected properly. | Ensure the Ethernet cable is firmly plugged in at both ends. | |
The Ethernet cable is defective. | Try using a different Ethernet cable. |
11.2 Error Messages and Warnings
Message | Description |
---|---|
Current compliance reached | The measured current is greater than the set current limit. |
Error communicating with system | The software is unable to connect to the system. |
No device holder detected | The device holder is not connected to the system properly. |
Sensor lid open | The lid of the device is not properly closed. |
Voltage increment cannot be zero | The voltage increment is set to 0 V. |
Start and end voltage cannot be equal | The start and end voltages of are set to the same voltage. |
No save directory or device label entered | The save directory and/or the device label fields are empty. |
Settings profile not found | The given settings profile does not exist or is open in other software. |
Error loading settings | There is a problem with the settings profile preventing it from being loaded. |
Error deleting profile | The given settings profile does not exist, the software does not have the necessary permissions to delete it, or it is open in other software. |
No data to save | There is no measurement data in memory to save to file. |
Error saving data | The software does not have the necessary permissions to access the given file path, or the file is already open in other software. |
Error saving settings | The software does not have the necessary permissions to access the given file path, or the file is already open in other software. |