FREE shipping to on qualifying orders when you spend or more, processed by Ossila BV. All prices ex. VAT. Qualifying orders ship free worldwide! Fast, secure, and backed by the Ossila guarantee. It looks like you are visiting from , click to shop in or change country. Orders to the EU are processed by our EU subsidiary.

It looks like you are using an unsupported browser. You can still place orders by emailing us on info@ossila.com, but you may experience issues browsing our website. Please consider upgrading to a modern browser for better security and an improved browsing experience.

Laminar Flow Hood Cleaning Procedure


Laminar flow hoods create a clean sterile environment for working with sensitive samples, such as biological cultures. Within the hood, HEPA filters ensure that most of the air entering your system is free from particles and contamination. However, there are still some ways by which your workspace can be compromised. Regular and thorough cleaning of your laminar flow hood allows you to reliably conduct your experiments without risk of contamination.

UV Sterilization


UV sterilization is a method used to kill microorganisms. Exposure to UV light alters the genes of harmful microbes, making them inactive. Therefore, UV lights are a common feature in equipment like PCR cabinets, many biological safety cabinets, and our laminar flow hoods. Some advanced clean rooms even use UV lights throughout the whole room.

Importantly, prolonged exposure to UV wavelengths can be harmful to humans. As a safety measure, workspaces should only be exposed to UV light to prepare them for sterile work. The light should always be turned off during workspace use. To prioritize user safety, Ossila laminar flow hoods are designed to automatically switch off the UV light when someone approaches. The UV light is only reactivated if motion has not been detected for 30 seconds. A warning light ensures you are always aware of the UV light status.

Preparing and Cleaning Your Laminar Flow Hood


Sources of Contamination

The primary sources of contamination in your laminar flow hood are:

  • Contamination that has accumulated on the work bench
  • Airborne contamination that has entered through the front panel
  • The surfaces of equipment and materials you bring into the laminar flow hood (bottles, vials, etc)

Cleaning Procedure

Cleaning Ossila Vertical Laminar Flow Hood

To clean your laminar flow hood, we recommend following specific steps:

  1. Spray the air with a disinfectant 10 minutes before beginning work. This neutralizes charges on colloidal particles, which causes them to fall out of the air.
  2. Immediately before working, spray surfaces with cleaning solution such as 70% ethanol solution or 80% isopropanol solution.
  3. Wipe all surfaces with a clean room tissue or microfibre cloth. Start with the cleanest area and progress to the dirtiest area to avoid spreading contamination. In a vertical laminar flow hood, clean the back wall first. In a horizontal laminar flow hood, clean the ceiling panel first.

  4. Clean the side walls by moving side to side, from top to bottom.
  5. Always clean the floor panel last. Repeat the same motion from the back to the front. During this stage, change the cloth or tissue often.
  6. Do not attempt to clean the filter panel.
Laminar flow Hood - Clean the back wall first
First, clean the back wall from top to bottom, with overlapping movements.
Laminar flow Hood - Clean the side walls
Then, clean the side walls using the same side-to-side motion.
Laminar flow hood - base cleaning - start at the back
At the base, start cleaning at the back of the flow hood
Laminar flow hood - base cleaning - front to back
Move towards the front of the workspace as you clean the base.

General Tips

  • You can use different cleaning solutions to clean your laminar flow hood, including IPA, ethanol, and distilled water. The solutions and steps you use should depend on your experiment. The general procedure is to use a 7:3 ethanol to water mix, or a 4:1 IPA to water mix.
  • Use several clean microfibre cloths or lint-free, non-shedding clean room tissue. Use a new cloth or tissue for each wall or whenever it looks dirty.
  • The entire hood should be cleaned daily. Clean the hood again if any contamination is suspected, after any spills, or before performing a media test.
  • Work surfaces should be cleaned before and after every use of the laminar flow hood.

Horizontal Laminar Flow Hood

  • Eco-Friendly
  • Fully Equipped
  • High Purity Air

Available Now! £2,500.00

Filter Contamination


If a HEPA filter becomes contaminated, it can quickly contaminate the whole workspace. This makes monitoring the saturation of your filter very important. It is also useful to be aware of any filter contamination risks.

As the filter traps more and more particles, it will become more saturated and less effective. Ossila laminar flow hoods will monitor and show the filter saturation on the display screen. This feature will notify you when the filter needs to be changed.

If the filter has become contaminated (causing contamination of your workspace), you can conduct a sterility experiment to check the integrity of your laminar flow hood.

Is Your Laminar Flow Hood Clean Enough?


Contamination is a risk in any environment. The likelihood of sample contamination occurring inside a laminar flow hood is greatly reduced, but it is not zero. You should periodically check the sterility of your laminar flow hood. This can be done using a simple sterility test. You can also use this technique to find the most sterile areas of your workspace, by placing multiple petri dishes at multiple points inside the hood.

For general testing, peptone agar plates can be used to check the sterility the workspace. Specific contaminants may require other media to thrive. For example, to test for fungal contamination you can use malt extract agar plates.

Agar plates in the laminar flow hood
Agar plates can be used to test the sterility of a laminar flow hood

Sterility Test

To conduct a sterility test:

  1. Expose the prepared media plates to the inside of the laminar flow hood for various time intervals. These intervals should be chosen depending on the type and length of experiment you are doing. 
  2. Incubate the petri dishes for 1–3 days
  3. Count the number of colonies that have grown

Exposure to the workspace begins once the protective lid of the media plate is removed and stops when the lid is placed back on.

Contamination Issues

The test indicates a sterile environment if there are no colonies after 1 hour of exposure. You should clean your laminar flow hood thoroughly if colonies grow following exposure. If this doesn't fix your contamination issues, you may need to:

  • Replace your HEPA filter
  • Replace any sterile materials you are using
  • Replace or clean any equipment you have brought into the laminar flow hood

If the test conditions change between the sterility test and your experiment, the results of the sterility test will be void. It is recommended that all required equipment is moved into the flow hood while sterility testing, to identify any contamination from external equipment.

References


Kawachi, M. et al. (2005) ‘Sterilization and sterile technique’, Algal culturing techniques. doi:10.1016/B978-012088426-1/50006-8