FREE shipping to on qualifying orders when you spend or more, processed by Ossila BV. All prices ex. VAT. Qualifying orders ship free worldwide! Fast, secure, and backed by the Ossila guarantee. It looks like you are visiting from , click to shop in or change country. Orders to the EU are processed by our EU subsidiary.

It looks like you are using an unsupported browser. You can still place orders by emailing us on info@ossila.com, but you may experience issues browsing our website. Please consider upgrading to a modern browser for better security and an improved browsing experience.

Linear Sweep Voltammetry: Introduction and Applications


Electrochemical cell with working, reference, and counter electrodes
An electrochemical cell with electrodes

Linear sweep voltammetry (LSV) is a simple electrochemical technique. The method is similar to cyclic voltammetry, but rather than linearly cycling over the potential range in both directions, linear sweep voltammetry involves only a single linear sweep from the lower potential limit to the upper potential limit. This is particularly useful for irreversible systems where a reverse sweep would not reveal any more information.

Like cyclic voltammetry, linear sweep voltammetry uses a three electrode system. This consists of a working electrode, counter electrode, and reference electrode.

The three electrodes are connected to a potentiostat and are placed in an electrochemical cell along with the solution of interest. The potentiostat controls the potential between the working electrode and reference electrode and measures the current at the counter electrode so that a plot of potential vs. current can be made. This shows the electrochemical response (oxidation or reduction) of the material in question. A peak or dip can be seen at the potential where oxidation or reduction begins.

Linear sweep voltammetry can be used to calculate the peak current, calculate the peak current potential, and calculate the half-peak current potential. For irreversible systems, linear sweep voltammetry is a simple and effective alternative to cyclic voltammetry.

Low Price Potentiostat

  • Cell and Electrodes Included
  • High Spec.
  • Intuitive Software

£2100 With Electrochemical Cell

Applications


Using linear sweep voltammetry, several key parameters can be observed which can be used diagnostically to determine thermodynamic reversibility of a material. These parameters include the peak current (ip), the potential at the peak current (Ep), and the potential at half the peak current (Ep/2) prior to the peak being reached.

These will be interpreted here for the system O + e- ⇌ R, with R initially absent (i.e. assuming no coupled reactions). These values can be used to calculate key electrochemical parameters at the reversible and quasireversible limits.

Note: the equations in this section assume linear diffusion. Most electrodes can be approximated as having linear diffusion on short time scales. The larger the electrode, and the shorter the time observed, the more the linear diffusion approximation applies.

The peak current

The peak current in amperes, ip, at 25 °C with bulk concentration O CO* in mol cm-3, electrode area A in cm2, the diffusion constant DO in cm2 s-1, and the scan rate v in V s-1 is given by:

Peak current in amperes equation

For a reversible system K is 1, and for an irreversible system:

Calculating K for an irreversible system

Graphs exist1, 2 to relate determine the degree of reversibility if α is known.

The peak current potential

The peak current is given by:

peak current potential equation

In the reversible system Ξ = 1.109 where as in the irreversible system:

Calculating E for an irreversible system

Graphs exist1,2 to relate determine the degree of reversibility if α is known.

The half-peak current potential

The half peak current potential is given by the following formula:

half peak current potentail formula

For a reversible system ∆ = 2.20 , whereas for irreversible system:

calculating delta for an irreversible system

Graphs exist1,2 to relate determine the degree of reversibility if α is known.

Low Price Potentiostat

  • Cell and Electrodes Included
  • High Spec.
  • Intuitive Software

£2100 With Electrochemical Cell

Keep Reading...


Ossila Potentiostat Getting Started with the Ossila Potentiostat

The Ossila Potentiostat has been designed to make it quick and easy to perform electrochemistry.

Read more...
Cyclic voltammetery graph Cyclic Voltammetry Basic Principles, Theory and Setup

Cyclic voltammetry is an electrochemical technique used to measure the current response of a redox active solution to a linearly cycled potential sweep.

Read more...

References


  1. A. J. Bard and L. Faulkner, Electrochemical Methods: Fundamentals and Applications, John Wiley & son, 2nd edn., 2001.
  2. H. Matsuda, Y. Ayabe, Z. Elektrochem., 1955, 59, 494-503.